Upgrade Notes
- Update your boxes. Most of the fixes in this release run on the box. Run
lager update --box <box-name>after you upgrade the CLI. lager wifi statusexits 1 when the box has no wireless interface. It used to print “Interface detection failed” and exit 0. A script that treats that exit as success must now handle the failure.lager wifi statusandlager wifi access-pointsrun on the box host over SSH. They used to run inside the container, which cannot see the host’s wireless interface.access-pointsnow scans the box’s own wireless interface by default, notwlan0.lager usbon a YKUSH hub fails when a port does not switch. The hub acknowledges a switch that it ignores, andenable,disable,toggleandcyclereported success. The driver now reads the port back. A script that relied on the false success now gets an error.
Features
- Yepkit YKUSH3 hubs are supported.
lager instrumentsandlager nets add-alldetect the YKUSH3 (04d8:f11b). Its three USB ports areusbnets forlager usb. Its three GPIO pins aregpionets forlager gpiandlager gpo. The hub’s switchable 5V output is not a net. - The IPEVO V4K PRO document camera (
1778:d00e) is detected as a webcam.lager instrumentslists it, andlager nets add <name> webcam ...accepts it. - A gated box’s gateway can see which command each request belongs to. Requests that carry
a
lager logintoken also send the command’s name, such assupply voltage, with the CLI version and an id for the command. The name never includes arguments. A box that is not behind a gateway receives none of these headers.
Bug Fixes
- The
/scopepage shows the scope’s real state.- A Rigol that loses only its capture card is no longer called unplugged. The page says that the live view stopped, and it says when the card is back.
- The page says when the box closes its connection after a long idle.
- The run badge shows a dash after a disconnect. It used to keep its last state, such as TRIG’D, until the next frame.
- The page says when a scope net has no channel nets, and it gives the
lager nets addcommand for each channel. - The stream request no longer returns paths on the box when a scope’s daemon fails.
- The
/scopepage draws a Rigol’s trace where the scope’s own screen does.display zoomnow magnifies the trace and shows the zoom bar.- A channel’s hardware offset now moves its trace. The volts read off the trace do not change.
- The box checks for a Rigol’s side menu again a minute after a check finds none. The menu no longer stays over the live view.
- A
stoptyped in the/scopepage’s console stops the measurement polling.startandstart singlein the console now follow the same rule as their buttons. lager scope <net> cursor move-aandmove-bmove from where the cursor is. On a Rigol, each call started from the center of the grid. Afterset-a --x -200,move-a --x 50now puts A at -150, not +50.lager scope <net> statussays “running” for a Rigol that the daemon opens while it runs. The command and the scope page said “stopped” after every daemon restart, power cycle or replug until you pressed Start.lager scope <net> measureprints a reading without float noise. A Rigol’s duty cycle now prints as49.987 %, not49.986999999999995 %. A Rigol’s refusal of an averaging count no longer names 65536 captures, which the CLI does not accept.- A Rigol MSO5000 bus decoder is refused before anything is sent. Creating a UART, I2C, SPI, CAN or FlexRay decoder failed partway through the setup with “Function not found”. It now raises an error that says the scope does not decode buses through lager.
- The scope daemon’s log says when the capture card stopped. Requests that wait at that moment are no longer told that the oscilloscope stopped answering.
- The scope daemon notices a PicoScope unplugged while it was idle. The next command failed with a bare driver error. The daemon now closes the unit when a 2000-series call fails and no PicoScope is on the bus. It opens the unit again when it returns.
- A PicoScope plugged back in gets its settings back. Channel enables, volts/div, coupling, probe, timebase and trigger returned to the driver’s defaults, so a page that pressed Start drew nothing. The same unit, by serial, now keeps its settings. A different unit starts from its defaults.
lager debug flashandlager debug eraseno longer call an unreachable box too old. With--erase-startor--erase-size, a/healthtimeout told you to update the box. The command now retries twice, then says that the box or its Stout gateway did not answer.lager diagnoseon a USB hub net reports an unreachable endpoint as unreachable. It used to say that the box does not serve/diagnose/usbhub, and it told you to update the box.lager usbsays that the box service restarts, not “cannot reach box”. When a hub does not open after it re-enumerates, the box service exits to recover. It now answers first, and it tells you to retry in a few seconds.- Python API calls on a multi-channel supply net act on that net’s channel. On a Rigol DP800
or Keysight E36000,
ovp(),ocp(),state(),clear_ovp()and similar calls throughNet.get()sometimes acted on another channel.lager supplywas not affected. lager supplycan set OVP and OCP back to a Rigol DP800’s factory limits. A DP800 ships with protection limits up to 110% of the channel’s rating, and the CLI refused any limit above the rating. The limit is now 110%. Voltage and current setpoints stay within the rating.wait_for_level()on a LabJack T7 GPIO net counts the stream setup against its timeout. A call ran about 30 ms past its timeout. The call still stops the stream before it raisesTimeoutError, which can take about 0.6 s.- A command stopped with
killreleases its box lock. A killed background command, such aslager debug <net> gdbserver, used to hold its lock until the lock’s TTL ran out. - The box’s authorized_keys sync survives the end of an install or update. The key-sync poller now runs in its own session. A signal to the caller’s process group no longer stops it.
- The box image carries the license notices of uldaq and the LabJack Exodriver. They are at
/usr/share/licenses/lager/vendor/.
Known Limitations
- The YKUSH3 GPIO pins are not fully verified on hardware. Their voltage level, and whether a written level is held, are not yet measured.
- Two IPEVO V4K PRO cameras on one box share an address. The camera reports no serial.
- I2C and SPI per-call overrides stay on the driver. An
overridesvalue such asfrequency_hzalso applies to later calls on the same net. Set the value that you want on each call. - A failed resistance set on an EA solar net does not raise an error. The call returns the resistance that the instrument still has. Compare the returned value with the value you set.